# FDA recall Z-1156-2021

> **Hamilton Co** · Class II · device recall initiated 2021-01-15.

## Product

HAMILTON HEATER SHAKER, 115 VAC/ 230 VAC, 750 VA, Power Consumption 41V/ 100W (max.), Relative Humidity 15% - 85% no condensation. REF 19903400

## Reason for recall

There is a risk that the device can start a fire if flammable liquids contact and enter the device.

## Distribution

US - AL, AR, AZ, CA, CO, CT, DC. DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA. MD, ME. MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI AND WY    OUS - AUSTRALIA, AUSTRIA, BRAZIL , CANADA    , CHILE, CHINA,    , COLOMBIA  , FINLAND   , GERMANY   , ITALY     , JAPAN     , JPN       , KOREA, SOUTH, PARAGUAY  , PERU      , PHL       , ROMANIA   , ROU       , SINGAPORE, SWITZERLAND       , SWITZERLAND        , THAILAND, TWN       , UNITED KINGDOM, AND URUGUAY

## Key facts

- **Recall number:** Z-1156-2021
- **Recalling firm:** Hamilton Co
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2021-01-15
- **Report date:** 2021-03-10

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Reno, NV, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1156-2021

## Citation

> AI Analytics. FDA recall Z-1156-2021. Retrieved 2026-08-11 from https://api.ai-analytics.org/recall/Z-1156-2021. Source: US FDA. Licensed CC0.

---