FDA recall Z-1158-2020

Roche Diagnostics Operations, Inc. · Class II · device

Product

cobas infinity core, Material No. 7154003001. The device is a calculator/data processing module, for clinical use.

Reason for recall

Alarm flags that have a "<" symbol in front of them are not displayed on the cobas infinity Validation screen. This issue affects only customers that are doing manual validation on the cobas infinity Validation screen.

Distribution

Distributed to customers in AK, AL, CA, GA, HI, IA, IL, KY, MA, MI, MN, MO, MS, MT, NJ, NY, OR, SC, and TN.

Key facts

Status
Terminated
Initiation date
2020-01-07
Report date
2020-02-19
Termination date
2020-10-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1158-2020