FDA recall Z-1159-2023

Cardinal Health 200, LLC · Class II · device

Product

Cardinal Health Presource Packs IV START KIT-Intended for IV Access CATALOG #: 01A4553E

Reason for recall

Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity

Distribution

US Nationwide distribution in the states of GA, PA.

Key facts

Status
Ongoing
Initiation date
2022-11-18
Report date
2023-03-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukegan, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1159-2023