FDA recall Z-1163-2025

Microbiologics Inc · Class II · device

Product

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test

Reason for recall

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

Distribution

Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.

Key facts

Status
Ongoing
Initiation date
2025-01-20
Report date
2025-02-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Cloud, MN, United States