FDA recall Z-1165-2020

Pacific Medical Group Inc. · Class II · device

Product

Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341

Reason for recall

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.

Distribution

Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.

Key facts

Status
Ongoing
Initiation date
2019-09-11
Report date
2020-02-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Clemente, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1165-2020