FDA recall Z-1183-2018

Quidel Corporation · Class II · device

Product

Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Reason for recall

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Distribution

US Distribution

Key facts

Status
Terminated
Initiation date
2017-12-15
Report date
2018-04-04
Termination date
2019-10-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source