FDA recall Z-1184-2018

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator is an in vitro diagnostic product for the calibration of follicle stimulating hormone (FSH), luteinizing hormone (LH), prolactin (PRL), and estradiol (E2) methods on the Dimension Vista System.

Reason for recall

An increase in the number of failed calibration events or negative bias with Quality Control (QC) and patient samples when using Dimension Vista LOCI 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay.

Distribution

Worldwide Distribution: US (nationwide) and countries of: Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Slovakia, Spain and Switzerland.

Key facts

Status
Terminated
Initiation date
2017-11-09
Report date
2018-04-04
Termination date
2020-06-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newark, DE, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1184-2018