FDA recall Z-1185-2021

Brainlab AG · Class II · device

Product

ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).

Reason for recall

In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.

Distribution

USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.

Key facts

Status
Terminated
Initiation date
2021-03-01
Report date
2021-03-17
Termination date
2023-03-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Munich, Germany

Primary source