# FDA recall Z-1185-2024

> **EMD Millipore Corporation** · Class II · device recall initiated 2024-01-26.

## Product

Schiff Reagent, Hotchkiss-McManus, packaged in an amber glass bottle with a black polypropylene cap.

## Reason for recall

This IVD is colorless. A customer complaint was received noting it appeared pink. A review of inventory confirmed additional stock appeared pink.

## Distribution

Domestic: AL, IL, MA, MN, NJ, PA, TX, VA.  Foreign: Canada.

## Key facts

- **Recall number:** Z-1185-2024
- **Recalling firm:** EMD Millipore Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-01-26
- **Report date:** 2024-03-06

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Burlington, MA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1185-2024

## Citation

> AI Analytics. FDA recall Z-1185-2024. Retrieved 2026-06-09 from https://api.ai-analytics.org/recall/Z-1185-2024. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*