FDA recall Z-1208-2025

INTELERAD MEDICAL SYSTEMS INCORPORATED · Class II · device

Product

IntelePACS (Image Fusion Module) - InteleViewer

Reason for recall

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia

Key facts

Status
Ongoing
Initiation date
2025-02-05
Report date
2025-03-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Montreal, N/A, Canada