FDA recall Z-1209-2025

Tyber Medical · Class II · device

Product

A.L.P.S. mvX - ANATOMIC LAT FIB PLATE 4H LT: Anatomic Lateral Fibula Plate, 4-hole, Left Intended for Fixation of fractures of the distal tibia Model/Catalog Number: 770708041

Reason for recall

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Distribution

Nationwide

Key facts

Status
Ongoing
Initiation date
2025-01-21
Report date
2025-03-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bethlehem, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1209-2025