# FDA recall Z-1210-2023

> **DiaSorin Molecular LLC** · Class II · device recall initiated 2023-01-30.

## Product

Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.

## Reason for recall

Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.

## Distribution

US Nationwide distribution including in the states of NJ, ME, ND, IL, MI, FL, GA, AZ, MA, CA, MO, SD, MN, PA, OH, VT, NE, WI, CO, DC, VA, TX, NH, IN, IA, MD, NC, AL, NY, KY, WV, RI, KS, NM, OK, WA, SC, CT, TN, NV, LA.

## Key facts

- **Recall number:** Z-1210-2023
- **Recalling firm:** DiaSorin Molecular LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-01-30
- **Report date:** 2023-03-15

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Cypress, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1210-2023

## Citation

> AI Analytics. FDA recall Z-1210-2023. Retrieved 2026-07-09 from https://api.ai-analytics.org/recall/Z-1210-2023. Source: US FDA. Licensed CC0.

---