FDA recall Z-1212-2023

LetsGetChecked Inc. · Class II · device

Product

LetsGetChecked Blood Sample Collection kit-In vitro diagnostic medical device specifically intended for the primary containment and preservation of blood samples.

Reason for recall

Users received sample collection kits that were assigned to another user. Incorrect shipping labels were applied to the collection kits. As a result, a number of users received sample collection kits that were assigned to another user resulting in a mismatch of user identification. The impact to users is that they may return the sample and obtain potential incorrect Hepatitis C Virus (HCV) results and/or the sample may be rejected due to mismatched demographics resulting in delayed testing and/or treatment.

Distribution

US Nationwide distribution in the state of Alabama.

Key facts

Status
Ongoing
Initiation date
2022-03-16
Report date
2023-03-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Astoria, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1212-2023