FDA recall Z-1213-2023

Baxter Healthcare Corporation · Class II · device

Product

Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis

Reason for recall

MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.

Distribution

US Nationwide distribution and Canada.

Key facts

Status
Ongoing
Initiation date
2023-02-01
Report date
2023-03-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1213-2023