FDA recall Z-1216-2018

Stryker Instruments Div. of Stryker Corporation · Class II · device

Product

The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are also helpful in determining the appropriate pedicle screw length.

Reason for recall

Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.

Distribution

International , US and US Govt.

Key facts

Status
Terminated
Initiation date
2017-10-16
Report date
2018-04-04
Termination date
2019-04-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portage, MI, United States