# FDA recall Z-1224-2023

> **St. Jude Medical** · Class II · device recall initiated 2023-02-07.

## Product

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

## Reason for recall

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

## Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.

## Key facts

- **Recall number:** Z-1224-2023
- **Recalling firm:** St. Jude Medical
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-02-07
- **Report date:** 2023-03-15

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Atlanta, GA, United States