FDA recall Z-1228-2019

OriGen Biomedical, Inc. · Class I · device

Product

ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV23F

Reason for recall

Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.

Distribution

US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile

Key facts

Status
Terminated
Initiation date
2017-01-17
Report date
2019-06-12
Termination date
2021-07-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Austin, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1228-2019