FDA recall Z-1230-2018

Zimmer Biomet, Inc. · Class II · device

Product

ACE Trochanteric Nail System STER TROC NAIL 180X11X140 STER TROC NAIL 125 DEG X 11MM STER TROC NAIL 130 DEG X 11MM STER TROC NAIL 135 DEG X 11MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.

Reason for recall

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

Distribution

USA (nationwide)

Key facts

Status
Terminated
Initiation date
2017-12-20
Report date
2018-04-04
Termination date
2020-04-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1230-2018