FDA recall Z-1230-2023

Datascope Corp. · Class II · device

Product

Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85

Reason for recall

There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.

Distribution

US Nationwide. Global Distribution.

Key facts

Status
Ongoing
Initiation date
2023-02-07
Report date
2023-03-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mahwah, NJ, United States