FDA recall Z-1233-2024

Ortho-Clinical Diagnostics, Inc. · Class II · device

Product

VITROS XT 7600 Integrated System, Catalog No. 6844461

Reason for recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2024-01-26
Report date
2024-03-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rochester, NY, United States