FDA recall Z-1235-2026

Beckman Coulter, Inc. · Class II · device

Product

CellMek SPS Sample Preparation System, REF: C44603, with software

Reason for recall

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Distribution

Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.

Key facts

Status
Ongoing
Initiation date
2025-12-23
Report date
2026-02-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Miami, FL, United States