FDA recall Z-1241-2025

GE Healthcare (China) Co., Ltd. · Class II · device

Product

Optima MR360, NMRI system

Reason for recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Distribution

Domestic US distribution nationwide. International distribution worldwide.

Key facts

Status
Ongoing
Initiation date
2025-01-31
Report date
2025-03-05
Voluntary/Mandated
Voluntary: Firm initiated