FDA recall Z-1249-2024

Canon Medical System, USA, INC. · Class II · device

Product

MRI system: Vantage Elan, Model: MRT-2020

Reason for recall

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Distribution

Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.

Key facts

Status
Ongoing
Initiation date
2024-01-29
Report date
2024-03-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tustin, CA, United States