# FDA recall Z-1256-2022

> **Steris Corporation** · Class II · device recall initiated 2022-04-22.

## Product

Reliance Synergy Washer/Disinfector

## Reason for recall

The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

## Distribution

Worldwide distribution - US Nationwide.

## Key facts

- **Recall number:** Z-1256-2022
- **Recalling firm:** Steris Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2022-04-22
- **Report date:** 2022-06-15
- **Termination date:** 2024-11-04

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Mentor, OH, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1256-2022

## Citation

> AI Analytics. FDA recall Z-1256-2022. Retrieved 2026-06-02 from https://api.ai-analytics.org/recall/Z-1256-2022. Source: US FDA. Licensed CC0.

---