# FDA recall Z-1266-2020

> **Theralase Inc.** · Class III · device recall initiated 2019-09-26.

## Product

TLC-2000 Therapeutic Medical Laser System

## Reason for recall

Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard

## Distribution

Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania

## Key facts

- **Recall number:** Z-1266-2020
- **Recalling firm:** Theralase Inc.
- **Classification:** Class III
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-09-26
- **Report date:** 2020-02-26
- **Termination date:** 2020-12-03

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** East York, Canada