FDA recall Z-1271-2021

Canon Medical System, USA, INC. · Class II · device

Product

Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000F Catheterization Table CAT-880B Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

Reason for recall

System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of Canada, China, Japan, Korea, and Singapore.

Key facts

Status
Ongoing
Initiation date
2020-12-08
Report date
2021-03-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tustin, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1271-2021