FDA recall Z-1272-2022

Smiths Medical ASD Inc. · Class I · device

Product

Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-402, 3500-414, 3500-415, 3500-500, 3500BC, 3500E, 3500G, 3500SD, 3500SD-500, 3500VX, 3500VX-306, 3500VX-414, 3500VX-415, 3500VX-500, 3500ZE, Software Versions 6.0.0, 6.0.1

Reason for recall

Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001

Distribution

Worldwide distribution. US nationwide including US Virgin Islands, Countries of: AE, AL, AT, AU, BE, BM, BN, BS, CA, CH, CL, CY, CZ, DE, DK, DO, ES, FI, FR, GB, HK, HU, ID, IE, IL, IN, IR, IT, KR, KW, LB, LK, MU, MX, MY, NL, NZ, OM, PA, PH, PK, PL, PT, QA, SA, SE, SG, TR, TT, VI, and VN.