FDA recall Z-1275-2023

Abbott Diabetes Care, Inc. · Class I · device

Product

FreeStyle Libre Reader, REF: 71951-01, 71952-01, 71953-01 a component of the FreeStyle Libre 2 Flash Glucose Monitoring System

Reason for recall

Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

Distribution

U.S. Nationwide.

Key facts

Status
Ongoing
Initiation date
2023-02-13
Report date
2023-04-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Alameda, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1275-2023