# FDA recall Z-1287-2023

> **Philips Respironics, Inc.** · Class I · device recall initiated 2023-03-01.

## Product

Trilogy Evo O2 and LifeVentEVO2 Portable Electric Ventilator, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, JP2100X16B, AU2100X15B, LA2100X15B, CA2100X12B, CN2100X17B, BR2100X18B, KR2100X15B, IA2100X15B, PP2100X10, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, BL2100X15B, GB2100X15B, EU2100X15B, EU2100X19, EE2100X15B, TR2100X15B, UDS2100X11B, RDS2100X11B, and SP2100X26B

## Reason for recall

The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

## Distribution

Worldwide distribution.   US Nationwide including Puerto Rico, and countries of AE, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GE, GT, HK, HR, HU, ID, IL, IE, IN, IT, JP, KW, LB, LT, MM, MQ, MX, MY, NG, NL, NO, NP, NZ, PE, PF, PH, PK, PL, PT, QA, RO, RS, RU, SA, SG, SI, SK, TH, TR, TW, UA, VN, and ZA.

## Key facts

- **Recall number:** Z-1287-2023
- **Recalling firm:** Philips Respironics, Inc.
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-03-01
- **Report date:** 2023-04-12

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Murrysville, PA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1287-2023

## Citation

> AI Analytics. FDA recall Z-1287-2023. Retrieved 2026-06-26 from https://api.ai-analytics.org/recall/Z-1287-2023. Source: US FDA. Licensed CC0.

---