FDA recall Z-1289-2023

Philips Respironics, Inc. · Class I · device

Product

Trilogy Evo Universal Portable Electric Ventilator, Model Number DS2000X11B

Reason for recall

The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

Distribution

Worldwide distribution. US Nationwide including Puerto Rico, and countries of AE, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GE, GT, HK, HR, HU, ID, IL, IE, IN, IT, JP, KW, LB, LT, MM, MQ, MX, MY, NG, NL, NO, NP, NZ, PE, PF, PH, PK, PL, PT, QA, RO, RS, RU, SA, SG, SI, SK, TH, TR, TW, UA, VN, and ZA.

Key facts

Status
Ongoing
Initiation date
2023-03-01
Report date
2023-04-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Murrysville, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1289-2023