FDA recall Z-1291-2025

DeRoyal Industries Inc · Class II · device

Product

DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement

Reason for recall

Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.

Distribution

U.S.: IL, MN, NC, PA, and VA O.U.S.: N/A

Key facts

Status
Ongoing
Initiation date
2025-01-31
Report date
2025-03-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Powell, TN, United States