FDA recall Z-1293-2023

Intuitive Surgical, Inc. · Class II · device

Product

da Vinci Xi Surgical System Model IS4000 da Vinci X Surgical System Model IS4200

Reason for recall

Due to Universal Surgical Manipulator (USM) instrument carriage being loose as a result of the linear rail component not being manufactured within specifications.

Distribution

U.S.: Cayman Islands, Dominican Republic, Puerto Rico, United States O.U.S.: Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Luxembourg, Martinique, Monaco, Netherlands, Norway, Reunion, Sweden, Switzerland, United Kingdom, India, Japan, South Korea, Taiwan, Brazil, Argentina, Australia, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Ecuador, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, Indonesia, Israel, Italy, Kuwait, Lebanon, Malaysia, Malta, Mexico, New Zealand, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Thailand, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam

Key facts

Status
Ongoing
Initiation date
2023-02-10
Report date
2023-04-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sunnyvale, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1293-2023