FDA recall Z-1294-2018

Seiler Instr. & Mfg. Co. Inc. · Class II · device

Product

Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Reason for recall

The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.

Distribution

Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.

Key facts

Status
Ongoing
Initiation date
2017-01-01
Report date
2018-04-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Louis, MO, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1294-2018