# FDA recall Z-1294-2019

> **Medica Corporation** · Class II · device recall initiated 2019-02-06.

## Product

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer   REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.

## Reason for recall

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

## Distribution

Worldwide distribution - US Nationwide and countries of   IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE  GUATEMALA, KENYA, IRAN, ESTONIA.

## Key facts

- **Recall number:** Z-1294-2019
- **Recalling firm:** Medica Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-02-06
- **Report date:** 2019-05-15
- **Termination date:** 2020-10-02

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Bedford, MA, United States