FDA recall Z-1294-2022

Siemens Medical Solutions USA, Inc · Class II · device

Product

Artis Pheno, Interventional Fluoroscopic X-Ray System, Model No. 10849000

Reason for recall

A video signal from one of the transceivers of the Large Display may be lost. In very rare cases, both transceivers of the Large Display could potentially fail which can lead to the loss of both video signals. This issue can sporadically affect the Large Display in the examination room as well as the displays in the control room.

Distribution

Domestic distribution nationwide. Foreign distribution to Australia Austria Belgium Brazil Canada China Colombia Denmark Finland France Germany Hungary Indonesia Iran Italy Japan Kuwait Mexico Monaco Netherlands Norway Poland Singapore Slovakia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom United States Vietnam

Key facts

Status
Ongoing
Initiation date
2022-06-10
Report date
2022-07-06
Voluntary/Mandated
Voluntary: Firm initiated