FDA recall Z-1297-2022

Atrium Medical Corporation · Class II · device

Product

Atrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the chest cavity or mediastinum. Product REF Number: 15028

Reason for recall

Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while the case and shipper labels both showed 21SEP2024 (2024-09-21), causing product being used for up to five (5) days past the true expiration date

Distribution

AZ IL KS LA MI NY OH OK TX VA

Key facts

Status
Ongoing
Initiation date
2022-04-21
Report date
2022-07-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Merrimack, NH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1297-2022