FDA recall Z-1298-2022

Haemonetics Corporation · Class II · device

Product

Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)

Reason for recall

When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.

Distribution

US Nationwide.

Key facts

Status
Terminated
Initiation date
2022-04-19
Report date
2022-07-06
Termination date
2024-03-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Boston, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1298-2022