# FDA recall Z-1300-2026

> **Maquet Cardiopulmonary Gmbh** · Class II · device recall initiated 2026-01-09.

## Product

Bubble Sensor for 3/8" x 3/32" tubing;  Model Number: 701055720;

## Reason for recall

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors  Ven. bubble sensor defective  or  Ven. bubble sensor disconnected  on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.

## Distribution

Worldwide Distribution:  US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam

## Key facts

- **Recall number:** Z-1300-2026
- **Recalling firm:** Maquet Cardiopulmonary Gmbh
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-01-09
- **Report date:** 2026-02-11

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Rastatt, N/A, Germany