FDA recall Z-1303-2024

Beckman Coulter, Inc. · Class II · device

Product

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Reason for recall

Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.

Distribution

Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.

Key facts

Status
Ongoing
Initiation date
2024-02-08
Report date
2024-03-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Chaska, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1303-2024