FDA recall Z-1309-2021

CareFusion 303, Inc. · Class I · device

Product

BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad; and Pump Module Door Assembly Replacement Kit with Pump Module Keypad, Parts: 49000239; 49000346; 49000438; 49000439.

Reason for recall

Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay

Distribution

Worldwide distribution. US nationwide and countries of: AE, IL, IN, SA, BE, QA, SG, TW, KW, ZA, PH, AU, HU, NZ

Key facts

Status
Ongoing
Initiation date
2021-03-03
Report date
2021-04-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1309-2021