FDA recall Z-1310-2019

Integra LifeSciences Corp. · Class I · device

Product

MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Valve w MoniTorr 10110 10-110 CSF Drainage System used with Pole Mount System MoniTorr 10140 10-140 CSF Drainage System w/Vinyl Measuring Strip MoniTorr 10150 10-150 CSF Drainage System Simple Bag and Line MoniTorr 10110A 10-110A CSF Drainage System w/o Manifold Stopcock MoniTorr SP0017 SP0017 Special EVD 10-110 w/o Y Site Latex Free Sites MoniTorr SP0042 SP0042 Special EVD 10-100 w/o Y Site MoniTorr SP0090 SP0090 Special EVD 10-110 w/o Y Site MoniTorr SP0164 SP0164 Special EVD 10-140 w/Y Site & Stopcock Reverse MoniTorr SP0236 SP0236 INS 1100 WITH NEEDLE AT Y-ACCESS AND NEEDLELESS SI Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, closed system for the drainage of cerebrospinal fluid (CSF) from ventricles of the brain or the lumbar subarachnoid space to a drainage bag. The system may

Reason for recall

The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.

Distribution

Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.

Key facts

Status
Terminated
Initiation date
2019-04-08
Report date
2019-05-29
Termination date
2021-01-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Plainsboro, NJ, United States

Primary source