FDA recall Z-1315-2023

Stryker Sustainability Solutions · Class II · device

Product

Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-235; Color Cuff Single Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-135; Color Cuff Dual Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-235; Color Cuff Single Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-135; Color Cuff Dual Port, Single Bladder 18 (Red), Quick Connect, REF: 5921-218-235;

Reason for recall

Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.

Distribution

US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC, VT, CT

Key facts

Status
Ongoing
Initiation date
2023-03-01
Report date
2023-04-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tempe, AZ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1315-2023