FDA recall Z-1317-2018

Smith & Nephew, Inc. · Class II · device

Product

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Reason for recall

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

Distribution

US, Puerto Rico, Malaysia

Key facts

Status
Terminated
Initiation date
2018-03-07
Report date
2018-04-18
Termination date
2018-10-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Memphis, TN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1317-2018