FDA recall Z-1318-2024

Think Surgical, Inc. · Class II · device

Product

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Reason for recall

There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.

Distribution

US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.

Key facts

Status
Ongoing
Initiation date
2024-02-12
Report date
2024-03-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Fremont, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1318-2024