FDA recall Z-1322-2023

Philips Respironics, Inc. · Class I · device

Product

Philips DreamStation Auto BiPAP and CPAP, Model Numbers UCDSX500S11, UDSX500S11, UDSX500S11F, UDSX700S11, UDSX700S11F, UFRX500S14. Used to deliver positive airway pressure therapy for the treatment of Obstructive Sleep Apnea.

Reason for recall

A limited number of remediated Philips DreamStation units may experience communication issues when connecting to the cloud-based care management application. Should this be the case, this may affect the ability to download prescription settings to the device or the device may have incorrect prescription settings for the patient.

Distribution

US Nationwide. France, Reunion.

Key facts

Status
Ongoing
Initiation date
2023-02-10
Report date
2023-04-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Murrysville, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1322-2023