FDA recall Z-1326-2020

Theralase Inc. · Class III · device

Product

TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.

Reason for recall

Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)

Distribution

Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.

Key facts

Status
Terminated
Initiation date
2019-09-26
Report date
2020-02-26
Termination date
2020-12-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
East York, Canada

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1326-2020