FDA recall Z-1326-2021

Philips North America, LLC · Class II · device

Product

Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481

Reason for recall

Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit

Distribution

International distribution in the country of China.

Key facts

Status
Terminated
Initiation date
2021-03-23
Report date
2021-04-07
Termination date
2023-09-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1326-2021