FDA recall Z-1329-2024

Baxter Healthcare Corporation · Class II · device

Product

HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery

Reason for recall

Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.

Distribution

US Nationwide distribution in the state of Florida.

Key facts

Status
Ongoing
Initiation date
2024-03-11
Report date
2024-03-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States