FDA recall Z-1337-2019

Terumo Cardiovascular Systems Corporation · Class II · device

Product

Terumo Advanced Perfusion System 1 Electronic Patient Gas System, Catalog # 801188, UDI 00886799000588 The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit.

Reason for recall

The gas flow rate output of the EPGS may be inaccurate due to a defect in the internal flowmeter.

Distribution

Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, Canada, Chile, Hong Kong, Indonesia, Japan, Mexico, Philippines, Singapore, South Korea, Taiwan, Thailand, and United Arab Emirates.

Key facts

Status
Terminated
Initiation date
2019-04-18
Report date
2019-05-22
Termination date
2020-05-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Ann Arbor, MI, United States

Primary source