FDA recall Z-1338-2023

Moberg Research, Inc. · Class II · device

Product

Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements Model: CNS-350

Reason for recall

Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC

Distribution

Worldwide distribution - US Nationwide and the countries of France, Germany, Italy, Sweden.

Key facts

Status
Ongoing
Initiation date
2023-02-20
Report date
2023-04-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Ambler, PA, United States